Excimer Laser System - Product Code LZS
원인
Erroneous treatment calculations - two software caused errors; in combination, in the wavescan system software, will result in an erroneous treatment calculation (overcorrection) in patients. (1) installation of software for a certain brand of computer (dedicated computing llc) can cause a software registry setting for an algorithm that identifies image reflection to be erroneously set in the "off.
조치
AMO sent an Action Required notification on May 30, 2007, to AMO Certified Physicians. The letter included a CD software update (0070-1721) to be installed by users that provided correct registry parameter settings. Upon completion of installation, a confirmation form will print out. Customers are requested to fax back confirmation to AMO (408-773-7055).
Worldwide Distribution --- USA including states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO,MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK,OR,PA, SC, TN, TX, UT, VA, WA, WI, and WV and countries of Canada and Mexico.
제품 설명
Advanced Medical Optics (AMO), VISX WaveScan WaveFront System (software). Part Number 0070-1478. Advanced Medical Optics, Santa Clara, CA 95051.