Device Recall XSYSTEMS Multiconstituent Controls for Abused Drug Assays 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Abbott Laboratories MPG 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    32540
  • 사례 위험등급
    Class 3
  • 사례 연번
    Z-1539-05
  • 사례 시작날짜
    2005-06-27
  • 사례 출판 날짜
    2005-09-14
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2006-01-18
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Drug Mixture Control Materials - Product Code DIF
  • 원인
    The xsystems multiconstituent controls accessory insert (commodity number 34-3067/r6) does not contain the correct control ranges for the axsym opiates assay.
  • 조치
    On 6/27/05, Abbott sent the customer letter via FedEx to all XSYSTEMS Multiconstituent Controls customers that received List Number 9687-12 lots 17270M100 and/or 18177M300 and/or AxSYM Opiates reagents (List Number 3B25-20) since June 01, 2004. The letters explained that the multiconstituent control ranges for morphine were not updated in the MCC Accesory insert at the same time they were updated in the AxSYM Opiates reagent package insert in 2004. The accounts were requested to review their current control ranges for the AxSYM Opiates assay, and verify that they are using either the control ranges in the AxSYM Opiates reagent package insert or control ranges established by their facility. They were also instructed to follow their facility''s normal policies and procedures for communicating with health care providers served by their laboratory. The accounts were requested to complete and return the enclosed reply sheet acknowledging the receipt of the letter and the dissemination of the letter to the laboratory manager/supervisor responsible for the assays.

Device

  • 모델명 / 제조번호(시리얼번호)
    lot numbers 17270M100 and 18177M300
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    Nationwide, including Puerto Rico, and internationally to Argentina, Australia, Barbados, Bermuda, Brazil, Canada, Cayman Island, Chile, China/Hong Kong, Columbia, Costa Rica, Dominican Republic, Ecuador, Guatemala, Germany, Honduras, Jamaica, Korea, Mexico, New Zealand, Panama, Peru, Singapore, St. Vincent, Thailand, Trinidad and Taiwan
  • 제품 설명
    XSYSTEMS AxSYM Multiconstituent Controls for Abused Drug Assays, list 9687-12; Abbott laboratories, Abbott Park, IL 60064 USA
  • Manufacturer

Manufacturer

  • 제조사 주소
    Abbott Laboratories MPG, 100 Abbott Park Rd, Abbott Park IL 60064-3502
  • Source
    USFDA