Device Recall ZeroP StarDrive Screwdriver Shaft 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Synthes Spine 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    51781
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1348-2009
  • 사례 시작날짜
    2009-02-10
  • 사례 출판 날짜
    2009-06-10
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2009-10-05
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    screwdriver instructions - Product Code ODP
  • 원인
    Tip breakage, due to inadequate instructions for use. the technique guide provided incorrect information related to use of the torque limiting attachment during screw insertion and final tightening.
  • 조치
    All affected Synthes Spine Sales Consultants were notified of the Labeling Correction via email on February 10, 2009. They were instructed to review changes with their surgeons. They were supplied with an electronic version of the corrected Technique Guide on March 11, 2009. All affected Synthes Sales Consultants and Customers were notified of this labeling change via distribution of the updated Technique Guide and a notification letter. This packet was sent via USPS certified mail on March 26, 2009. To ensure a broad and accurate distribution of notification, affected consultants and customers are defined as having been shipped Zero-P 3.0 mm Titanium Cervical Spine Locking Screws. These devices are used only with the Zero-P system.

Device

  • 모델명 / 제조번호(시리얼번호)
    Catalog # 03.617.902 - All lots
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    Yes
  • 유통
    Distributed Nationwide and Canada.
  • 제품 설명
    Synthes Zero-P StarDrive Screwdriver Shaft, T8, self-retaining, quick coupling, for use in spinal fusion surgery. Catalog # 03.617.902 Technique Guide.