DOLPHIN INFLATION DEVICE CALIBER INFLATION DEVICE 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 PEROUSE MEDICAL 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    77100
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1891-2017
  • 사례 시작날짜
    2017-04-05
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Syringe, balloon inflation - Product Code MAV
  • 원인
    Complaints regarding broken blister.
  • 조치
    Consignees were notified of the recall on April 5, 2017. They were informed of the issue and asked to take the following steps: 1. Inspect your stock to determine if you still have in your ownership devices among which the references and batch codes are listed in front page; 2. Control the potential damage of the blister according to the pictures above and the instructions mentioned on the labeling and the Instruction For Use: any damaged and/or cracked blister must be immediately discarded; 3. Use the compliant products without risk, after control. Indeed, the primary packaging will not deteriorate in time. If, to date, there is no crack in the blister, then the integrity of the product is not compromised; 4- Do not use the non-compliant products; 5. Complete and sign the attached reply form, then return it within 5 working days following the reception of this letter, and keep a copy; 6. Contact customer service for the modalities of return, replacement and associated support at +33 (0)4.72.39.74.13 or by fax at +33 (0}3.44.08.17.67 or by email at iblayon@vvgon.com (CC mhpourriere@vygon.com); 7. Communicate the safety information to any person concerned in your company and to all end customers to which the product were transferred.

Device

  • 모델명 / 제조번호(시리얼번호)
    1 5045606 - 15055430 - 15065340 - 15065685 -15065744- 15065858- 15065970- 15075121 - 15075265 - 15075411- 15075821 - 1 5075851 - 15075942 - 15085001 - 15085075 -15085378- 15085435 -15085515- 15085555 -15105052- 15125214- 16015267- 16025034- 16025640
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    US: AZ (shipped through Belgium to USA), MN, TN
  • 제품 설명
    DOLPHIN INFLATION DEVICE CALIBER INFLATION DEVICE; Catalog Number: CL3030- 018SNA- 018SND 0185NA 0185ND; INTERNAL CODE NUMBER: 0185TR- 0185NA- 018SND 0185NA 0185ND
  • Manufacturer

Manufacturer

  • 제조사 주소
    PEROUSE MEDICAL, ROUTE DU MANOIR, Ivry Le Temple France
  • 제조사 모회사 (2017)
  • Source
    USFDA