U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Incorrect dicom mapping of the exported collimator or couch angles from ergo, which would lead to incorrect rotation of the collimator or couch when using a mlc device for planning.
조치
Elekta sent an Urgent Important Field Safety Notification letter dated September, 2017. The letter identified affected product, problem and actions to be taken. Customers were instructed that a copy of the Important Field Safety Notice is to be kept with the most current labeling and all personnel working with the product should be made aware of the content of the letter. The customers were asked to complete the verification tests and return the Important Field Safety Notice Acknowledgment form to their local Elekta representative as soon as possible but no later than 30 days from receipt. For questions contact your local Elekta Care Support Center.
Worldwide Distribution - US Nationwide in the states of: VA, CA, TX, MO, NY, WA, NM, MA, FL, CT, OK, SC, MS, NY, GA, NC and the countries of: AUSTRALIA, BANGLADESH, BRAZIL, BULGARIA, CANADA, COLOMBIA, CZECH REPUBLIC, ECUADOR, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, IRELAND, ITALY, JAPAN, JORDAN, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, PERU, PHILIPPINES, PORTUGAL, RUSSIA, SAUDI ARABIA, SINGAPORE, SOUTH KOREA, SPAIN, TAIWAN, THAILAND, TURKEY, VENEZUELA, and VIETNAM
제품 설명
Elekta ERGO++ || Product Usage: || ERGO is often used for stereotactic treatments.