EMPOWR PS KNEE(TM) e 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Encore Medical, Lp 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    78193
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0042-2018
  • 사례 시작날짜
    2017-09-15
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code JWH
  • 원인
    It was reported that an implant labeled as 343-13-708 was packaged with implant 343-19-707.
  • 조치
    The firm initiated the recall by email on 09/05/2017. The letter identified the affected product, problem and actions to be taken. The consignee was directed to notify anyone to whom the product was transferred and to return any in-stock product.

Device

  • 모델명 / 제조번호(시리얼번호)
    Lot Number 055U1003
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    Yes
  • 유통
    US Nationwide Distribution
  • 제품 설명
    EMPOWR PS KNEE(TM) e+, REF 343-19-707, SIZE 7, 19mm, QTY 01, GREY, STERILE || Product Usage The EMPOWR PS Inserts are manufactured from ultra high molecular weight polyethylene (UHMWPE) infused with pure liquid pharmaceutical grade alpha-tocopheral and then cross-linked to conform to ASTM F2695. This insert is posterior stabilized and is intended to more closely match the kinematics of the knee, allowing some rotation along the medial condyle and increased congruency along the lateral condyle.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Encore Medical, Lp, 9800 Metric Blvd, Austin TX 78758-5445
  • 제조사 모회사 (2017)
  • Source
    USFDA