Entroy Pool Lift 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Arjo, Inc. dba ArjoHuntleigh 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    59588
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-3070-2011
  • 사례 시작날짜
    2011-08-04
  • 사례 출판 날짜
    2011-08-24
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2011-11-28
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Lift, patient, non ac-powered - Product Code FSA
  • 원인
    The entroy pool lift boom may rapidly lower un-commanded to the bottom of its stroke, with a potential for injury to the resident or caregiver.
  • 조치
    ArjoHuntleigh sent an Urgent Field Correction Notice I.D. 110701AHE to the affected end user via UPS 2nd Day on August 4, 2011. The notice informed the account of the potential for lift boom to rapidly lower un-commanded to the bottom of its stroke. The account was requested to immediately cease use of the affected lift, and disconnect the power to the lift to prevent further use until corrected. The account was requested to complete and sign the Customer Response Form, and return it to ArjoHuntleigh to schedule a service call to correct the lift. Questions were directed to the ArjoHuntleigh at 1-800-323-1245, ext 54485.

Device

  • 모델명 / 제조번호(시리얼번호)
    Model GAB1000-01-US, serial number SEE0326028  Rest of the world: Models GAB1000-01-xx, GAB1000-11-xx and GAC100x, all serial numbers
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution - USA including CT
  • 제품 설명
    Entroy Pool Lift; a patient lift used in conjunction with a chair or stretcher to lift them in/out of a swimming/therapy pool; Arjo Hospital Equipment AB, Eslov, Sweden; Models GAB1000-01-xx, GAB1000-11-xx and GAC100x || Entroy is a pool lift that docks to either a chair or a stretcher and aids disabled people getting in and out of a pool. This equipment is intended for disabled persons who need to transfer in and out of a therapy and swimming pool under supervision of skilled and trained personnel in accordance with the Entroy Operating and Product Care Instructions. All other use must be avoided. Active/semi-active persons can and should use the Entroy chair (with or without chassis) while bedridden persons without trunk stability can and should be transferred with the Entroy stretcher with chassis.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Arjo, Inc. dba ArjoHuntleigh, 2349 W Lake St, Addison IL 60101-6183
  • 제조사 모회사 (2017)
  • Source
    USFDA