U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
데이터 추가 비고
Introducer, catheter - Product Code DYB
원인
St. jude medical is conducting a voluntary recall of smj batch # 4641580 of the fast-cath trio" hemostasis introducer, catalog number 406308. the inner diameter of the 10f dilator is too small, such that the guidewire cannot advance through the device which may result in a procedural delay while another dilator of the desired size is obtained.
조치
St. Jude Medical sent an Urgent Medical Device Recall Notice dated October 14, 2014, to all affected consignees. The letter described the problem and the product affected by the recall. Advised consignees to discontinue us of the device. A St. Jude Medical sales representative were to contact the consignee in the near future to facilitate removal and reconciliation of the recalled device. Customers with questions were instructed to contact their St. Jude Medical sales Representative.
For questions regarding this recall call 651-756-2295.
Nationwide Distribution including FL, IN, ME, MA, NY, NY, OH, RI,TX, and WV.
제품 설명
St Jude Medical, 10F Fast-Cath Trio" Hemostasis Introducer, REF # 406308, Sterile EO. || The Fast-Cath Trio" Hemostasis Introducer consists of the following components: Port hemostasis adapter (3), Sheath (10F Cath-Lock"), dilator (10F) and 0.038 guidewire. See product labeling for diagram of contents.
“We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.