fluoroscopic imaging system - Product Code MQB
원인
Incorrect dose data: ge healthcare has found that after 6 days and 4 hours without performing a system reset or a system reboot (turning it off/on), the displayed dose data on these systems (innova 3100 / innova 4100 with bolus option) may be underestimated by up to 50%. in order to avoid this issue, it is necessary to perform a system reset or reboot every day. ge healthcare is working on a c.
조치
GE sent all customers with affected units a letter on 11/15/07 titled Urgent Medical Device Correction. In the US this was sent by a traceable means (either US Postal Service with receipt acknowledgement (registered mail) or equivalent courier service such as FEDEX overnight Delivery). The letter was sent to 3 titles within the affected accounts; Hospital Administrator, Manager of Radiology/Cardiology and Radiologist/ Cardiologist. A correction letter identified a short term recommendation to perform a system reset or reboot every day and stated that a software correction will be installed at their site at no cost.