Gyrus ACMI Round Cutting Burrs 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Gyrus ACMI, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    79548
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1337-2018
  • 사례 시작날짜
    2017-08-03
  • 사례 출판 날짜
    2018-03-20
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Bur, ear, nose and throat - Product Code EQJ
  • 원인
    "chattering" when cutting burrs are used in bone.
  • 조치
    As a result, OSTA will issue a Field Safety Notice (FSN) to all customers that received variable exposure burrs of 5mm or greater diameter. The FSN will inform them of the concern and will advise to exercise due caution if using the affected device in proximity to sensitive or vital anatomical structures. On September 6, 2017 the firm sent a URGENT: MEDICAL DEVICE FIELD SAFETY NOTICE to the attention of the Surgical Risk Management Department for the following products: MBUR5060FRCV - Burr, 5mm Round Cutting, Variable Exposure, Standard Length, MBUR6060FRCV - Burr, 6mm Round Cutting, Variable Exposure, Standard Length and MBUR7060FRCV - Burr, 7mm Round Cutting, Variable Exposure, Standard Length for All product with an expiration date prior to August 3, 2017 Description of the problem: Olympus has become aware of an issue that requires your attention. This letter pertains to the Olympus Round Variable Cutting Burrs (Burrs) referenced above. Our records indicate that you may have purchased affected burrs. We are in receipt of a number of non-injury-related complaints that has made Olympus aware of a possibility that the burr may chatter when drilling bone. Although this chatter phenomenon can occur with any type or brand of cutting burr, depending on technique and bone density, Olympus has become aware that the larger diameter variable exposure burrs listed above can exhibit more chatter than other similar devices in some situations, particularly when they are extended to longer lengths. After extensive testing, Olympus has determined that our variable exposure burrs of less than 5mm diameter, and all fixed exposure burr offerings, are not affected by this phenomenon. Because of this, Olympus has elected to issue this field safety notice (FSN). Please refer to the enclosure with this letter that cautions of the possibility of the chatter phenomenon in the larger variable offerings. The FSN also recommends that surgeons exercise due caut

Device

  • 모델명 / 제조번호(시리얼번호)
    Catalog No. (UDI-DI): MBUR5060FRCV (00821925036529), MBUR6060FRCV (00821925036536), and MBUR7060FRCV (00821925036543) ALL LOTS MANUFACTURED BEFORE AUGUST 3, 2017
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    NM, MI, and TX EU, Australia, Japan, and Hong Kong
  • 제품 설명
    Gyrus ACMI Round Cutting Burrs - to be used with the Diego Elite system
  • Manufacturer

Manufacturer

  • 제조사 주소
    Gyrus ACMI, Inc., 2925 Appling Rd, Bartlett TN 38133-3901
  • 제조사 모회사 (2017)
  • Source
    USFDA