Hamilton C2 Ventilator 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Hamilton Medical, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    66408
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0170-2014
  • 사례 출판 날짜
    2013-11-06
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2018-03-29
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Ventilator, continuous, facility use - Product Code CBK
  • 원인
    Manufacturer defined limits for improperly functioning loudspeakers of the hamilton c2 alarm system were exceeded.
  • 조치
    Hamilton Medical sent an Medical Device Safety Alert and Corrective Action letter dated September 27, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The corrective action does not require any action by the operator. In the case of an alarm situation as described in the letter, users should follow the instruction for use (Section 8 - responding to alarms in the operator manual) and have the ventilator serviced by an authorized service technician. An addendum to the Operators Manual as a result of the field correction was provided to customers with the Safety Alert Letter. Manufacture will develop and provision a test protocol for the service tech which test the proper functioning of the loudspeaker during the annual service inspection. We appreciate your support in this matter and sincerely regret any inconveniences that this action may cause you. We consider this action as necessary to ensure that our customers receive only safe and effective products with high quality. Further questions please call (775) 426-6331 Ext. 210

Device

  • 모델명 / 제조번호(시리얼번호)
    Hamilton-C2; Serial numbers from 1000-7100.
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Nationwide distribution: USA including Puerto Rico and the states of IL, MT, IN, ND, AZ, MT, KY, NC, OH, MA, MO, VA, CA, NV, TX, WA, NY, GA, SC,MN, WY, PA, ND, UT, NE, CO and MI.
  • 제품 설명
    HAMILTON-C2 Ventilator: || Manufactured by Hamilton Medical AG || CH-7402 Bonaduz, Switzerland; || Imported and distributed by Hamilton Medical, Inc. Reno, NV 89502 || The HAMILTON C2 ventilator is intended to provide positive pressure ventilatory support to adults, pediatrics, infants and neonates depending on its configuration. Intended areas of use: In the intensive care ward or in the recovery room; During transfer of ventilated patients within the hospital.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Hamilton Medical, Inc., 4990 Energy Way, Reno NV 89502-4123
  • 제조사 모회사 (2017)