birthing bed - Product Code KNC
원인
The cable on the auxiliary outlet may become pinched, which may result in an electrical short with melting of the plastic transformer and the emission of smoke.
조치
Customers were notified via letter dated 8/18/06 and instructed to either correct the problem themselves with the instructions provided or to contact the firm requesting a visit from the recalling firm to correct their beds.
Nationwide, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, France, French Polynesia, Germany, Guam, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Republic of Korea, Kuwait, Lebanon, Libya, Malaysia, Mexico, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Sait Kitts and Nevis, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, U.S. Virgin Islands and Venezuela.
제품 설명
Hill-Rom Affinity Three Birthing Bed, Model P3700.