Analyzer, Chemistry (Photometric, Discrete), For Clinical Use - Product Code JJE
원인
Version 2.6 software was released however it will not properly handle adjustment slopes.
조치
Recall Technical Bulletins were sent out on 6/13/2003 to all Immulite 2000 users instructing them to not install the provided version 2.6 software.
Software was distributed to Immulite 2000 users throughout the nation. These units are found in hospitals and clinics. Software was also distributed to the firm''s service representatives.
제품 설명
Immulite 2000. Immunoassay system. Software version 2.6 was to be uploaded on systems running 2.5b software.