Impaction fixture 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Encore Medical, Lp 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    72229
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0180-2016
  • 사례 시작날짜
    2015-09-14
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2016-11-04
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Impactor - Product Code HWA
  • 원인
    During the assembly, the impaction forces caused the polymer, black acetal copolymer from the impaction fixture to wear off on the lateral surface of the humeral stem titanium plasma spray coating.
  • 조치
    DJO Global sent an Urgent Field Safety Notice dated September 15, 2015, to all affected consignees. The notice identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed become familiar with the new techniques and notify all of their applicable users of the RSP Monoblock and Turon stems of these revised methods. When each user has been notified of these new techniques, they should complete the form that was provided with the notification indicating that they have been informed of these changes and understand the new methods. Consignees were asked to complete and return forms. Consignees with questions should call 512-834-6255.

Device

  • 모델명 / 제조번호(시리얼번호)
    Lot # 52748L02A
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    Nationwide Distribution
  • 제품 설명
    Reverse Shoulder Prosthesis Stem Impaction Fixture REF 804-03-053 || The Turon and RSP Impaction Fixtures are used in the operating room on the back table to assemble the shoulder systems for implant. The Turon Fixture is used to connect the Humeral Head to the Humeral Neck, then subsequently to assemble the Humeral Stem to the Head/Neck construct. The RSP Fixture is used to assemble the UHMWPE Socket Inserts to both the RSP Socket Shells and Monoblock Stems.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Encore Medical, Lp, 9800 Metric Blvd, Austin TX 78758-5445
  • 제조사 모회사 (2017)
  • Source
    USFDA