INSIGNIA Ultra and NEXUS Ultra 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Guidant Corporation 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    33491
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0188-06
  • 사례 시작날짜
    2005-09-22
  • 사례 출판 날짜
    2005-09-24
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2006-12-27
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Implantable Pacemaker Pulse-Generator - Product Code DXY
  • 원인
    A failure mode occurring at a low rate, has been identified within the insignia and nexus families of implantable pacemakers. device behaviors which may be observed include: intermittent or permanent loss of pacing output without warning, intermittent or permanent loss of telemetry and appearance of a reset warning message upon interrogation.
  • 조치
    Important Medical Device Safety Information & Corrective Action , dated 09/22/05 was sent to all affected physicians. This letter explained the failure mode, what devices are affected and gave recommendations. A press release was issued on 09/22/05.

Device

  • 모델명 / 제조번호(시리얼번호)
    INSIGNIA Ultra SR (model 1190) serial no. 889963 - 891468; NEXUS Ultra SR (model 1390) serial no. 100007 - 100034. INSIGNIA Ultra DR (model 1290) serial no. 100678 - 105229; (model 1291) serial no. 100279 - 100913; NEXUS Ultra DR (model 1490) serial no. 100007, 100008, 100009, 100013, 100014, 100015, 100016, 100017, 100018; (model 1491) serial no. 100007, 100008, 100009, 100010, 100013, 100015, 100016, 100017, 100018.
  • 의료기기 분류등급
  • 의료기기 등급
    3
  • 이식된 장치?
    Yes
  • 유통
    Nationwide and Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Cyprus, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, Guadeloupe, Hong Kong, Ireland, Israel, Italy, Jordan, Korea, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Monaco, Netherlands, Netherlands Antilles, New Zealand, Norway, Panama, Polynesia, Portugal, Qatar, Reunion, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom and Venezuela
  • 제품 설명
    Guidant INSIGNIA I Ultra family of pacemakers includes the following: SR (models1190); and DR (models1290, 1291). Intermedics NEXUS I Ultra family of pacemakers includes the following: SR (model 1390) and DR (model 1490, 1491). The INSIGNIA I Ultra pacemakers are multiprogrammable pacemakers from Guidant. The NEXUS I Ultra pacemakers are multiprogrammable pacemakers from Intermedics. The family consists of both dual-chamber and single-chamber models, offering adaptive-rate therapy and providing various levels of therapeutic and diagnostic functionality. These pacemakers include ventricular Automatic Capture which automatically measures the ventricular pacing threshold and adjusts the pacing output to 0.5 V above the measured threshold. Two sensors are available: these adapt the pacing rate to the patient''s changing metabolic demand. Minute ventilation responds to changes in respiration, and the accelerometer responds to patient activity (motion). INSIGNIA and NEXUS I Ultra models can use either the accelerometer or minute ventilation sensor, or a blend of both accelerometer and minute ventilation. || Sterilized with gaseous ethylene oxide. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. or Intermedics, Inc., 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Guidant Corporation, 4100 Hamline Ave N, Saint Paul MN 55112-5798
  • Source
    USFDA