Integra Ruggles Surgical Instrument ACF Distraction Screws 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Integra LifeSciences Corp. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    50112
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0854-2009
  • 사례 시작날짜
    2008-10-22
  • 사례 출판 날짜
    2009-01-23
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2009-07-15
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Orthopedic Manual Surgical Instrument - Product Code LXH
  • 원인
    A packaging error had occurred where 16 mm screws were packaged and labeled as 12 mm screws and vice versa.
  • 조치
    Urgent: Product Recall Notification letters were sent via Fed Ex on October 22, 2008. The letter asked customers to quarantine any of the affected stock on hand and contact their local Integra Sales Representative for further assistance and for arrangement of product returns. Customers were to fill out the attached "Recall Acknowledgement and Return Form" indicating whether or not they possess any of the product identified in the letter. Additional questions can be directed to Christin Roach at 609-936-5413. A Urgent: Product Recall Second Notification dated 1/5/2009 was sent to customers providing them with additional information on the packaging error and included immediate and/or product used during surgery. This notification also requested that all surgeons be notified who were or may have been provided with any of the screws affected by this recall.

Device

  • 모델명 / 제조번호(시리얼번호)
    lot code 326081; and lot code 327081.
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    Nationwide Distribution
  • 제품 설명
    Integra Ruggles Surgical Instrument ACF Distraction Screws || Sterile, Rx only || Catalog number R6357A (12 mm), Catalog number R6397A (16 mm), || Integra LifeSciences Corp., 311 Enterprise Drive, Plainsboro, NJ 08536
  • Manufacturer