Kodak DirectView DR 7100 System, Catalog #133-6049, a digital radiology system. The major system components are: operator console; digital imaging capture device; tilting table with floating top; ceiling-mounted x-ray tube; x-ray generator timing and distribution unit. 의 리콜
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
데이터 추가 비고
System, X-Ray, Stationary - Product Code KPR
원인
Defective component in overhead tube support balancer may cause the suspended load to fall.
조치
2/20/2004: Field representatives notified the customer in China that the unit should not be used until the OTS Balancer component can be repaired/replaced.
Kodak DirectView DR 7100 System, Catalog #133-6049, a digital radiology system. The major system components are: operator console; digital imaging capture device; tilting table with floating top; ceiling-mounted x-ray tube; x-ray generator timing and distribution unit.