LifeScan 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Lifescan Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    31640
  • 사례 위험등급
    Class 1
  • 사례 연번
    Z-0823-05
  • 사례 시작날짜
    2005-04-11
  • 사례 출판 날짜
    2005-05-17
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2007-01-24
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    System, Test, Blood Glucose, Over The Counter - Product Code NBW
  • 원인
    User may inadvertently change the units of measure from mg/dl to mmol/l and the blood glucose results could be misinterpreted. this may lead to under treatment and potential for hyperglycemia.
  • 조치
    On 4/11/05, the firm issued letters via Federal Express to all its direct consignees, informing them of the affected products and providing instructions on the correction. Owner's booklet: Firm contacted all accounts by letters of 11/21/03, 2/4/2004, 2/17/04 and 2/9-10/2004 providing important information regarding the owner's booklet.

Device

  • 모델명 / 제조번호(시리얼번호)
    ALL CODES: All lot numbers and serial numbers
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Distribution of the product is worldwide. ***Distribution in the United States was between January 15, 2001 for OneTouch Ultra Blood Glucose Meter; April 13, 1998 for OneTouch FastTake Blood Glucose Meter and November 30, 2001 for InDuo Blood Glucose Meter.*** The total number of quantities shipped to consignees is approx. 1.8 Million. Of these, Approx. 282,000 were shipped to Pharmacies and Healthcare Professionals (including Diabetes Educators, Primary Care Physicians, Endocrinologists), Approx. 800 were shipped to Mail Order companies that distribute OneTouch ¿ FastTake¿ and OneTouch ¿ Ultra¿ products, and Approx. 130,000 were shipped to E-mail End User Notifications to LifeScan database of OneTouch¿ Ultra¿, OneTouch¿ FastTake¿ and InDuo¿ product users. Ultra, InDuo, FastTake Meters Global Distribution include: Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Caribbean, CEE, Chile, China, Columbia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Georgia, Germany, Greece, Hong Kong, India, Indonesia, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Kuwait, Latin America, Latin America, Lebanon, Malaysia, Malta, MEA, Mexico, Netherlands, Norway, Oman, Pasean, Peru, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia/Gulf Countries, Singapore, Slovakia, Slovenia, South Africa, Spain, Sub-Saharan Africa, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, United States, Uruguay, Venezuela, Vietnam, and Yemen. Owner's booklet: The recall product was distributed in the U.S. only. LifeScan manufactured and distributed OneTouch Ultra System Kits with the affected OneTouch Ultra Meter Owner's Booklets between March 31, 2003 and October 3, 2003.
  • 제품 설명
    LifeScan Brand InDuo Blood Glucose Meter
  • Manufacturer

Manufacturer

  • 제조사 주소
    Lifescan Inc, 1000 Gibraltar Dr, Milpitas CA 95035-6301
  • Source
    USFDA