Medtronic GEM DR 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Medtronic Inc. Cardiac Rhythm Managment 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    28798
  • 사례 위험등급
    Class 1
  • 사례 연번
    Z-0812-04
  • 사례 시작날짜
    2004-04-02
  • 사례 출판 날짜
    2004-07-20
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2005-11-29
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Implantable Cardioverter Defibrillator (Non-Crt) - Product Code LWS
  • 원인
    Some of the devices with suspect capacitors (high voltage capacitors associated with prior notifications in 1999 and in 2000) have had unexpected charge circuit time-outs or charge circuit inactive conditions as the battery voltage nears the level for elective replacement of the devices.
  • 조치
    A letter, dated April 2004, was sent on April 5, 2004 to physicians following patients who had the defibrillators implanted. The letter gave recommendations to physicians for selective replacement of some of the implanted devices and for avoiding the problem. Physicians are being visited by sales representatives to verify notification.

Device

  • 모델명 / 제조번호(시리얼번호)
    Manufacturing Dates from May 1997 to August 1998; and  Use Before Dates from November 1998 to February 2000.
  • 의료기기 분류등급
  • 의료기기 등급
    3
  • 이식된 장치?
    Yes
  • 유통
    Nationwide in the United States and worldwide.
  • 제품 설명
    Medtronic GEM DR Implantable Cardioverter Defibrillators, Model 7271