U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
데이터 추가 비고
Pump, Infusion, Insulin - Product Code LZG
원인
Several complaints of pumps exhibiting a47 alarms were received. investigation revealed the alarms occurred when the pump attempted to display text that exceeded the maximum display length of the pump screen. the investigation revealed this only occurs when the pump is programmed to display information in spanish.