Dose and mu are incorrect when ct images are viewed from the head, and, when using multiple prescriptions with forced densities.
조치
The firm issued an URGENT IMPORTANT FIELD SAFETY NOTICE on 09/15/2015. The letter identified the affected device, as well as the reason for the recall, and clinical impact. Customers were informed that both issues would be resolved in Monaco Patch Release 5.10.01. The attached acknowledgement form should be completed, signed, and returned to a local Elekta Office or representative within 30 days. Questions regarding the notice should be directed to a local Elekta office.
Worldwide Distribution -- US, including the states of CA, IN, MI, MO, TX and WI; and, the countries of Australia, Canada, Germany, Greece, India, Netherlands, New Zealand, Turkey, and the UK.
제품 설명
Monaco Radiation Treatment Planning System. || Used to make treatment plans for patients with prescriptions for external beam radiation therapy.