Novocastra Thyroid Transcription Factor1 (liquid) 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Leica Microsystems, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    56406
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-2499-2010
  • 사례 시작날짜
    2010-08-02
  • 사례 출판 날짜
    2010-09-25
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2012-01-29
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Immunohistochemistry Reagents And Kits - Product Code NJT
  • 원인
    In combination with a sensitive polymer detection system, the ttf-1 antibodies may stain additional tumor types other than is currently indicated on the instructions for use.
  • 조치
    Leica Microsystems sent Medical Device Field Correction letters dated 7/28/10 to the affected customers via first class mail on the same date, informing them that in combination with a sensitive polymer detection system, the TTF-1 antibodies may stain additional tumor types other than is currently indicated on the Instructions for Use. Currently they state that lung and thyroid tumors may stain, but Leicas investigation further characterized the expression of the TTF-1 antigen and it may also be present in other tumors, e.g. thymomas. Customers who have received the affected lot numbers of the three products were requested to refer to the new Instructions for Use to review the amendments made to the Results Expected via the following links: http://bsd.leica-microsystems.com/pdfs/products/ttf-1-ce.pdf; http://bsd.leica-microsystems.com/pdfs/products/ttf-1-l-ce.pdf; and http://bsd.leica-microsystems.com/pdfs/products/pa0364.pdf; The accounts were asked to complete the attached Acknowledgement Form, acknowledging the receipt and understanding of the letter, listing the lot numbers of the units at their facility and fax it back to Leica at 1-847-607-3147. Additional questions should be addressed to the firm's office of Regulatory Affairs and Quality Assurance.

Device

  • 모델명 / 제조번호(시리얼번호)
    Catalog #NCL-L-TTF-1, lot numbers L136434, L136437, L136443, L136446, 6000568 and 6002638.
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    Nationwide and Canada
  • 제품 설명
    Liquid Mouse Monoclonal Antibody Thyroid Transcription Factor-1; 1 mL liquid tissue culture supernatant in a 1 mL vial; an in vitro diagnostic reagent for the qualitative identification by light microscopy of thyroid transcription factor-1 molecules in paraffin-embedded tissue by immunohistochemical staining; Leica Biosystems Newcastle Ltd., Balliol Business Park West, Benton Lane, Newcastle Upon Tyne NE12 8EW, United Kingdom; Catalog No.: NCL-L-TTF-1
  • Manufacturer

Manufacturer

  • 제조사 주소
    Leica Microsystems, Inc., 2345 Waukegan Rd, Bannockburn IL 60015-1515
  • Source
    USFDA