NucliSENS Lysis Buffer 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 BioMerieux SA 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    78387
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0086-2018
  • 사례 시작날짜
    2017-10-18
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    General purpose reagent - Product Code PPM
  • 원인
    Elute may become colored due to the residual presence of heme group origination from hemoglobin present in whole blood samples as well as dried spot (dbs) samples. this can potentially cause the inhibition of pcr reactions and result in an uninterpretable test result. if using an internal control (ic) in the extraction process the ic could potentially be inhibited as well and invalidate test results.
  • 조치
    US customers were notified on 10/18/2017 via letter using 2-day Fed Ex shipment. Instructions included notifying all appropriate personnel in the laboratory or to customers if product was further distributed, refrain from using the affected product, and arrange for replacement of affected product by completing and returning the response form. It also instructed customers to contact and discuss any concerns regarding previously reported results with the appropriate Laboratory Medical Director to determine an appropriate course of action.

Device

  • 모델명 / 제조번호(시리얼번호)
    REF 200292, Lot 17022802, Exp Date 01/28/2019  Extended lot for RES 76675
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    Distributed domestically to MD, CO, and OH.
  • 제품 설명
    NucliSENS Lysis Buffer || Extended lot for RES 76675