Reamer - Product Code HTO
원인
Investigation determined that units supplied were missing the 4x21 degree helix angle.
조치
On September 17, 2014, "URGENT MEDICAL DEVICE RECALL NOTICE" was sent to the consignee Biomet Global Supply Chain Center (GSCC) by email and FedEx. GSCC notified all of the customers within Europe, the Middle East, and Asia that received the affected product. The recall notification included a description of the reason for recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call (574) 372-1570.
International Distribution in the countries of Netherlands, Europe, the Middle East, and Asia.
제품 설명
Compress Face Reamer || Product Usage: || Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement, tumor resections, revision of previously failed total joint arthroplasty, and trauma.
“The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.