Permobil Chairman 2K powered wheelchair 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Permobile Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    25733
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0834-03
  • 사례 시작날짜
    2003-03-07
  • 사례 출판 날짜
    2003-05-15
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2008-06-25
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Wheelchair, Powered - Product Code ITI
  • 원인
    One of the wheelchair motors needed a grounding strap to ensure that the motors are at the same voltage potential as the chassis of the wheelchair.
  • 조치
    Permobil Inc has notified its field representatives of the voluntary recall and these representatives have been provided with lists of each Chairman 2K serial number and the Dealer to which the wheelchair was sold. This list also includes the name of the individual for whom the wheelchair was specified at the time of order from Permobil Inc, where applicable. The Permobil field representatives have been trained to perform the upgrade and have been provided the necessary tools to complete the task. They will also train and supervise the Dealer''s technicians who will effect the upgrade at the Dealer level, as the field representatives will rely upon their Dealer network to complete the upgrades in a timely fashion.

Device

  • 모델명 / 제조번호(시리얼번호)
    Serial numbers: 1100155 to 1102472 and 4100019 through 4100225.
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Nationwide, Canada and Puerto Rico
  • 제품 설명
    Permobil Chairman 2K powered wheelchair
  • Manufacturer

Manufacturer

  • 제조사 주소
    Permobile Inc., 6961 Eastgate Blvd, Lebanon TN 370906005
  • Source
    USFDA