Test, system, immunoassay, lipoprotein-associated phospholipase a2 - Product Code NOE
원인
Inaccurate results-- some samples may return with falsely elevated results potentially as a result of an interfering substance. some patients may be subjected to additional lifestyle changes or intensified pharmaceutical therapy; such as statins, as a result of the falsely elevated result.
조치
Notification letter issued to consignees on 10/29/2009. Customers were instructed to not report PLAC test results for samples that are greater than 360 ng/mL, and pursue confirmatory testing by contacting DiaDexus. Customers with questions please call the company technical support line toll free at 1-877-752-2837 for assistance.
UPDATE: On April 30, the firm suspended marketing of the PLAC TIA method, as the limited range no longer covered the clinical data/reference range data cited in the product insert. The firm sent out a letter notifying customers on May 10, 2010. The firm set up a sample send out program, so that customers can send samples to labs running the ELISA method PLAC test. Arrangements were made so that labs could be enabled to run a CLIA complex assay, for customer credit programs for un-used products, and a fax back response form for response tracking.
Product was distributed to 123 consignees throughout the US.
제품 설명
PLAC Test Reagent Kit, immuno-turbidimetric method, Part Number 10-112, 10-0113, and 10-115 Advia test kit, manufactured by DiaDexus, South San Francisco, CA.