U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
ICU Medical, Inc. acquired from Pfizer in early 2017, Hospira Infusion Systems, the portion of Hospira dedicated to develop infusion pumps.
데이터 추가 비고
Infusion Pump - Product Code FRN
원인
Fire/shock hazard-- a component used in the device may cause sparks/flashes and poses a fire and shock hazard.
조치
Firm initiated its recall on 08/14/2009. A nationwide press release was issued to AP, and an Urgent: Device Field Correction letter to consignees was sent via UPS. The letter recommended that customers check their infusion pumps and spare parts in their inventory to determine if the specified AC Power Cords are in use at their facility; and, to inspect the affected cords for bent or cracked prongs and bridge connection, burnt plastic, or excessive wear and tear. Also, it was recommended that customers immediately discontinue use of the affected cords exhibiting the described characteristics and contact their local Hospira representative for assistance. The letter stated that customers may continue to use and monitor cords that are not exhibiting the described characteristics until replacement cords are sent to their facility. Hospira will contact customers to schedule replacement. Customers are to complete and return the attached Reply Form. Question should be directed to Hospira Customer Service at 1-800-241-4002.
Plum XLD, list numbers 11859-04-01, 11859-04-03, 11859-04-05, 11859-04-07, 11859-04-09, 11859-04-11, 11859-04-13, 11859-04-51, 12570-04-01 (veterinary); AC cord manufactured by Electri-Cord, Westfield PA, Infusion pumps manufactured by Hospira, Morgan Hill, CA.