PURE OXYGEN CONCENTRATOR, CHAD Drive 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Inovo, Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    78883
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0407-2018
  • 사례 시작날짜
    2017-10-09
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Generator, oxygen, portable - Product Code CAW
  • 원인
    Post-market surveillance for pure stationary oxygen concentrators model chsooo and chsooos demonstrate an aggregate complaint/service rate greater than 35% for low oxygen purity (output oxygen below the published specification).
  • 조치
    The firm initiated their recall by letter on 10/09/2017. The letter stated the following: "Actions to be Taken: Based on our distribution records, your firm has received one or more Pure Stationary Oxygen Concentrators. We respectfully request that you take the following actions: 1. Contact all patients and customers that currently have Pure Stationary Oxygen Concentrators in their homes or facilities and inform them of this recall. Retrieve the oxygen concentrators from their homes or facilities. 2. Destroy all Pure Stationary Oxygen Concentrator(s) in accordance with the following steps: a. Assure that the unit is not plugged into an electrical outlet. b. Cut the power cord of each oxygen concentrator. c. Take a digital photo of each oxygen concentrator that depicts both the power cord and the serial number of the depicted unit. Note that the serial number is located on a label on the back of the unit. d. Dispose of the oxygen concentrator in accordance with local ordinances. e. Record the serial number of each unit that is destroyed, sign and date the enclosed Recall Return Response Card. 3. Email or fax the completed response form and photos documenting destruction to: Email: lnovo4209@stericycle.com Fax: 1-877-497-2559 4. Sales of the Pure Stationary Oxygen Concentrator have been discontinued. Pure Stationary Oxygen Concentrators which have been destroyed will be replaced with a DeVilbiss 525DS Oxygen Concentrator in order to minimize any interruption in patient's supplemental oxygen therapy. 5. Pure Concentrators that have been returned to lnovo for repair will not be returned and will be replaced with a DeVilbiss 525DS Oxygen Concentrator. Other Information: Should you require any additional information about this recall action, please call 1-888-679-5132, Monday-Friday 8am-Spm Eastern."

Device

  • 모델명 / 제조번호(시리얼번호)
    UDI #(s): 00814470020013 (model CHSOOO) and 00814470020006 (model CHSOOOS): All serial numbers
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    worldwide
  • 제품 설명
    PURE OXYGEN CONCENTRATOR, CHAD Drive, Models CH5000 and CH5000S
  • Manufacturer

Manufacturer

  • 제조사 주소
    Inovo, Inc, 401 Leonard Blvd N, Lehigh Acres FL 33971-6302
  • 제조사 모회사 (2017)
  • Source
    USFDA