Test, Time, Prothrombin - Product Code GJS
원인
The isi labeling for simplastin¿ htf on the coag-a-mate¿ max and thrombolyzer instrument platforms is not correct. the simplastin htf isi assignment was mis-assigned and the published values are higher than expected for the listed instrument platforms.
조치
Consignees were notified by letter on/about March 7, 2005.
Product numbers 259846, 259847, Lot numbers: 161849, 161800 and 161798. Recall Expanded (03/23/2005) to include these additional lot numbers: 161655, 161710, 161741, 161738, 161742, 1621739, 161743, 161740, 161764, 161763, 161797 and 161799.