When sonocalc imt 3.0 software is used with the sonosite 180plus ultrasound system, the software miscalculates the intima-media-thickness, a measurement of risk for cardiovascular or cerebrovascular events. underestimates percentage of stenosis, false negative result.
조치
On October 26, 2005, the company sent their customers a User Guide Addendum that told customers how to verify that the image calibration data was correct.
Nationwide, Kuwait, Mexico, South Africa, Chile, Spain, Japan, Puerto Rico, Australia. A list of 155 cases in which the software was sent out lists medical facilities, physicans, and SonoSite Distributors.
제품 설명
SonoCalc IMT 3.0, a windows based software used in conjunction with images from high-resolution ultrasound systems (MicroMaxx, TITAN, and SonoSite180PLUS). This software generates a report with the patients intima media thickness (IMT) based on average thickness of the carotid artery.