Spacelabs Medical 1400 MHz Telemetry System,. Model 91341-09. Telemetry System consists of Model 91341-09 transmitter, 90478-1CST receiver and 90478-1CSV receiver. 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Spacelabs Medical Incorporated 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    35391
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1202-06
  • 사례 시작날짜
    2006-05-05
  • 사례 출판 날짜
    2006-07-06
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2010-12-27
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Detector And Alarm, Arrhythmia - Product Code DSI
  • 원인
    Potential for patient waveforms to move to an open receiver module or to an occupied receiver causing the intermittent inappropriate monitoring of both patients for several minutes.
  • 조치
    On May 5, 2006 a letter was issued to all consignees, via return receipt requested. The letter advised consignees of the issue and provided some ways to mitigate the risk. Anyone experiencing the anomaly is to document the failure and report it to the firm. A field service representative will visit each consignee to optimize the antenna system to reduce intermodulations. After determining the root cause and validating a fix, the firm issued a second letter dated August 21, 2006 to their consignees. The letter advised consignees of three issues that needed 'improvement' and stated a service engieneer will schedule a time to visit and install the improvements.

Device

  • 모델명 / 제조번호(시리얼번호)
    All systems - each major component has its own serial number
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    The systems were distributed to 22 hospitals/medical centers nationwide and to 2 Spacelabs personnel. There is no international distribution.
  • 제품 설명
    Spacelabs Medical 1400 MHz Telemetry System,. Model 91341-09. Telemetry System consists of Model 91341-09 transmitter, 90478-1CST receiver and 90478-1CSV receiver.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Spacelabs Medical Incorporated, 5150 220th Ave Se, Issaquah WA 98029-6834
  • Source
    USFDA