Accelerator, linear, medical - Product Code IYE
원인
When the stealth chamber is used as a reference detector in relative dosimetry, deviation in the measured pdds at shallow depths and geometric shadowing of measured profiles may appear.
조치
IBA Dosimetry sent an Field Safety Notice dated April 21, 2017. The letter informed the consignee of the issue and provided the following instruction: If a Stealth Chamber was used during commissioning of Linear Accelerator and Treatment Planning System, the consignee was asked to consider referenced deviations for larger field sizes (>20 x 20 cm) and higher energies (>60MV). They were also instructed to recheck the influence of the PDD deviation and insufficient clearance for 40 x 40 profile on their beam model and its clinical relevancy. For further questions please call (901) 268-6430.
Contact information
Should you have questions or require additional information, please contact the Customer Service Team at::
South and Middle America, USA, and Canada :
IBA Dosimetry America
Bartlett, TN, USA
US Service@.iba-group .com
Phone: +1 901 386-2242
Fax: +1 901 382-9453
Asia Pacific, A Australia and
New Zealand:
IBA China, Dosimetry Dept.
Beijing, China
Service APAC@.iba-group .com
Phone: +86 10 8080 9107
Fax: + 86 10 80809298
All other countries:
IBA Dosimetry GmbH, Service Dept.
Schwarzenbruck , Germany
service@.iba-group .com
Phone: +49 9128 607-38
Fax: +49 9128 607-26
Worldwide Distribution - USA (nationwide) and to the countries of : Argentina, Australia, Austria, Canada, Cyprus, France, Germany, Hong Kong, Ireland, Islamic Republic of, Israel, Italy, Japan, Kingdom of Saudi Arabia, Kuwait, New Zealand, People's Republic of China, Philippines, Russia, Slovakia, South Africa, South Korea, Spain, Sweden, Taiwan, R.O.C., and United Kingdom.
제품 설명
Large Stealth Chamber, Catalog Numbers: DS50-001-XL-T, DS50-002-XL-T, DS50-003-XL-T, DS50-102-XL-T, DS50-103-XL-T