STERRAD 100NX Cassettes 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Advanced Sterilization Products 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    66317
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0044-2014
  • 사례 시작날짜
    2013-09-18
  • 사례 출판 날짜
    2013-10-22
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2014-01-13
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Sterilizer, chemical - Product Code MLR
  • 원인
    Advanced sterilization products (asp) is recalling the three lots (13f049, 13f050, 13f051) sterrad 100nx cassettes because asp has determined that the leak indicator on the cassettes is effective for only 12 months of the cassette's labeled 15-month shelf-life.
  • 조치
    Advanced Sterilization Products (ASP) sent an Urgent Medical Device Recall letter on September 18, 2013, to all affected customers. The letter informed the customers of the problems identified and the actions to be taken. Customers were instructed to examine their inventory and return all affected cassettes using the enclosed prepaid UPS return lave to Stericycle. Customers were asked to count their inventory and record the data on the enclosed Business Reply Card and Packing Slip. Customers were asked to provide the notice to anyone in their facility that needs to be informed. Customers were also asked to maintain a copy of the notice with the affected product. Customers with questions or suspected problems were instructed to contact Stericycle directly at 1-877-257-7119. For questions regarding this recall call 949-453-6400.

Device

  • 모델명 / 제조번호(시리얼번호)
    Lot# 13F050, 13F051, 13F049
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Nationwide Distribution
  • 제품 설명
    STERRAD 100NX Cassettes || The STERRAD Sterilization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.