Passer, Wire, Orthopedic - Product Code HXI
원인
The wire appears to be soft and the wire bent during insertion. this causes kinks in which the reamer could not pass over the bent/kinked wire.
조치
The firm distributed a recall letter to their EBI Fracture Sales Representatives and their EBI Trauma Distributors on 3/16/2005.
These devices are distributed nationwide to sales representatives who provide the product directly to the hospital/physician at the time of the surgery. The product was also distributed to Italy and Spain. There are no government accounts.
제품 설명
Threaded Tip GD Pin 1/8in x 9in Pk 5 Sterile found in Uniflex Profile Tibial Nail P/N 329640.