VITEK 2 AST N203 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Biomerieux Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    77026
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-2389-2017
  • 사례 시작날짜
    2017-04-20
  • 사례 출판 날짜
    2017-05-22
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    unknown device name - Product Code N/A
  • 원인
    The integrity of the product container may be compromised thereby exposing product to moisture which could lead to antibiotic degradation (loss of potency).
  • 조치
    The firm posted their recall notification on their customer website on 4/20/2017. The firm sent the recall notification letter to their customers via FedEx on 4/21/2017 for delivery 4/24/2017. A follow-up recall notice will be issued on 5/08/2017 which identified 3 additional lots impacted by the recall.

Device

  • 모델명 / 제조번호(시리얼번호)
    5730188403 02/Jun/18
  • 유통
    Nationwide distribution in the U.S, including military and government sites. The product was also distributed to Algeria, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia - Herzegovina, Brazil, Brunei, Bulgaria, Burkina Faso, Cambodia, Cameroun, Canada, Chad, Chile, China, Colombia, Congo, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Dutch Antilles, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Gabon, Gambia, Georgia, Germany, Greece, Grenada, Guatemala, Guinea, Haiti, Honduras, Hungary, India, Indonesia, Iraq, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Uganda, Kuwait, Laos, Latvia, Lebanon, Lithuania, Malaysia, Mali, Malta, Mexico, Mongolia, Morocco, Mozambique, Myanmar, Netherlands, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, and Zimbabwe.
  • 제품 설명
    VITEK¿ 2 Gram Negative Susceptibility card (AST-N203), REF 412864, 20 cards per carton.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Biomerieux Inc, 595 Anglum Rd, Hazelwood MO 63042-2320
  • 제조사 모회사 (2017)
  • Source
    USFDA