VITEK 2 Gram Negative Susceptibility Card, ASTN183 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Biomerieux Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    58135
  • 사례 위험등급
    Class 1
  • 사례 연번
    Z-2049-2011
  • 사례 시작날짜
    2011-04-04
  • 사례 출판 날짜
    2011-04-27
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2012-06-01
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    System, test, automated, antimicrobial susceptibility, short incubation - Product Code LON
  • 원인
    The piperacillin/tazobactam drug on the antimicrobial susceptibility cards is providing false susceptible and false resistant results for e. coli, as well as an additional 7 organisms; and the same drug is providing false resistant results for k. pneumoniae.
  • 조치
    BioMerieux sent an Urgent Product Correction Notice (EXPANDED RECALL), VITEK 2 Piperacillin/Tazobactam Test, Document #9300786," letter dated March 30, 2011, via Fed EX on April 4, 2011. to all affected customers. The letter describes the issues, how the issues impact the test results, describes the product changes that are going to take place, and provides instructions on required actions to be taken. The product changes will involve the eventual changeover to new cards that will be created which do not contain Piperacillin/Tazobactam (TZP2). The new cards will begin shipping by July 16, 2011, but in the meantime, the cards currently in possession of the customers can be used with specific required actions that need to be followed. Until the TZP redevelopment is completed and a resolution can be implemented, BioMerieux requires their customers to immediately take the following actions: The customers will be required to suppress the TZP2 resistant and susceptible results for Escherichia coli, Morganella morganii, Proteus mirabilis, Proteus vulgaris, Providencia rettgeri, Providencia stuartii, Pseudomonas aeruginosa, and Salmonella enterica and use an alternate method of testing. In addition, the customer is to suppress the TZP2 resistant result only for Klebsiella pneumonia and use an alternate method of testing. The letter provides instructions for configuring the VITEK 2 software to suppress the results from reporting. The instructions contained in the recall notification are to be applied to all current and future product lots that contain TZP2 until further notice. The customer is to complete the enclosed acknowledgement form and return it via FAX to acknowledge receipt of the notification and is requested to file the correction notice with their VITEK 2 Systems permanent records according to their internal policy or procedure. For questions or additional information customers were instructed to call 1-800-682-2666, option 3. In addit

Device

  • 모델명 / 제조번호(시리얼번호)
    All lot numbers
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    There was no U.S. distribution as the cards were for export only to Canada and Sweden.
  • 제품 설명
    VITEK 2 Gram Negative Susceptibility Card, AST-N183, REF 411 021, For In Vitro Diagnostic Use, 20 cards/box. The firm names on the label are bioMerieux, Inc., Durham, NC, and bioMerieux sa, Marcy-l'Etoile France. || The cards are intended for use with the Vitek 2 systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic/anaerobic gram-negative bacilli to antimicrobial agents.
  • Manufacturer

Manufacturer