Weck Hemolok XL Polymer Ligating Clips 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Teleflex Medical 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    35316
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1098-06
  • 사례 시작날짜
    2006-04-12
  • 사례 출판 날짜
    2006-06-09
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2007-11-14
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Polymer Ligating Clips - Product Code FZP
  • 원인
    The hem-o-lok ligating clips may become dislodged following ligation of the renal artery after laparoscopic donor nephrectomy. hem-o-lok ligating clips are now contraindicated for use in ligating the renal artery during laparascopic nephrectomies in living donor patients.
  • 조치
    Teleflex Medical sent Important Product Safety Information letters dated 4/18/06 to all of their customers who have purchased Hem-o-lok Ligating Clips from Teleflex. The letters informed the healthcare providers that Teleflex has been made aware of incidents in which the ligating clips (sizes L and XL) were reported to have become dislodged following ligation of the renal artery after laparoscopic donor nephrectomy. Although none of the incidents appear to have involved any defect in or malfunction of the ligating clips, misapplication of the Hem-o-lok clips during such laparascopic procedures may not immediately be apparent and can have serious, even life-threatening consequences post-operatively. Therefore, Teleflex has added the following contraindication to the Instructions for Use accompanying the Hem-o-lok Polymer Ligating Clips: 'Contraindications': Hem-o-lok ligating clips are contraindicated for use in ligating the renal artery during laparoscopic nephrectomies in living donor patients. 'Caution': The clip must be latched to ensure proper ligation of the vessel or tissue. Inspect the ligation site after application to ensure proper closure of the clip. Security of the closure should be confirmed after ligation. The Hem-o-lok Polymer Ligating Clip is not designed for use as a tissue marker. Weck recommends that more than one clip be used to ligate the renal artery in procedures other than laparoscopic donor nephrectomy (see Contraindication, above). Application of more than one clip to all other vessels should be left to the surgeon's judgment.' Any questions concerning the letter were directed to Teleflex Medical at 800-334-9751, option #7, then option #5. Teleflex sent follow-up ¿Important Product Correction Notice¿ letters dated 8/2/06 to all of their Hem-o-lok Ligating Clips customers via registered mail on 8/7/06. The letters provide clarification of the 4/18/06 letter, reinforcing that the clips are contraindicated for use in ligating t

Device

  • 모델명 / 제조번호(시리얼번호)
    catalog number 544250, all lots
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    Yes
  • 유통
    Nationwide
  • 제품 설명
    Weck Hem-o-lok XL Polymer Ligating Clips; Rx, sterile, extra large size, non-absorbable polymer ligation clips; 6 clips per cartridge, 14 cartridges per sales unit, 12 sales units per case; Mfg. by Weck Closure Systems, Research Triangle Park, NC 27709; REF 544250
  • Manufacturer

Manufacturer

  • 제조사 주소
    Teleflex Medical, 2345 Waukegan Rd Ste 120, Bannockburn IL 60015-1580
  • Source
    USFDA