Xlumena brand NAVIX Access Device 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Xlumena, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    64965
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-2167-2013
  • 사례 시작날짜
    2013-04-17
  • 사례 출판 날짜
    2013-09-09
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2013-09-23
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Laparoscope, general & plastic surgery - Product Code GCJ
  • 원인
    A complaint investigation has found that product fractures can occur at the distal end of the catheter under load.
  • 조치
    Xlumena sent an Urgent Medical Device Recall letter dated April 17, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discontinue use and segregate the affected product. Customers with questions were instructed to contact dhasker@xlumena.com. For question regarding this recall call 650-961-9901.

Device

  • 모델명 / 제조번호(시리얼번호)
    All lots are involved are being recalled, including FGS00048, FGS0005, FGS00058, FGS00059, FGS00069, FGS00073, FGS00081, FGS00086, FGS00091
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution - USA including AZ, CA, FL, GA, IN, NY,SC, WA, PA, IL, CO, and internationally to Netherlands, Germany, Spain, Denmark, and Switzerland.
  • 제품 설명
    Xlumena brand NAVIX Access Device, Model/Catalogue Number: NVX-10-03, Part Number: FG-00666, Product is manufactured and distributed by Xlumena, Inc., Mountain View, CA || The NAVIX Access Device is a multi-lumen balloon catheter with a trocar, anchor balloon, and dilation balloon that creates and dilates a tract across tissue planes. The NAVIX Access Device is indicated for use as an accessory to cannulate the transgastric or transduodenal wall and into a pancreatic pseudocyst, when it is visibly bulging into the gastrointestinal tract during endoscopic procedures.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Xlumena, Inc., 453 Ravendale Dr Ste H, Mountain View CA 94043-5221
  • 제조사 모회사 (2017)
  • 제조사 의견
    “We take a patient-first approach to assessing the applicability of every recall and communicate to regulatory bodies in all geographies where the recalled device is sold,” Boston Scientific said in a statement to ICIJ. “We have coordinated several recalls across many countries in a timely manner,” the company said, adding that it complies with all national laws, which can often vary and require different processes for reporting information or taking action on recalls. The company said it uses a rigorous and uniform process to take action on recalls and that “when we initiate a field action (e.g. recall, safety alert), every customer who has received an affected product receives a communication that includes a letter for the physician.”
  • Source
    USFDA