Device Recall Philips HeartStart

  • 모델명 / 제조번호(시리얼번호)
    Serial Numbers in the following ranges: USO1100100 to USD1101095, US11201096 to USD1203968, US11303969 to USD1309471, US11409472 to US61414022
  • 의료기기 분류등급
  • 의료기기 등급
    3
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution - US Nationwide including PUERTO RICO, Canada, ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BANGLADESH, BELGIUM, BOLIVIA, BULGARIA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA (local name: Hrvatska), CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICA, DOMINICAN REPUBLIC, EGYPT, EL SALVADOR, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GABON, GERMANY, GREECE, GUATEMALA, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAN (ISLAMIC REPUBLIC OF), IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, Jordan, KENYA, KOREA, REPUBLIC OF, KUWAIT, LATVIA, LEBANON, LIBERIA, LIBYAN ARAB JAMAHIRIYA, LITHUANIA, MACAU, MALAYSIA, MALTA, MAURITIUS, MEXICO, MOROCCO, MYANMAR (Burma), NAMIBIA, NEPAL, NETHERLANDS, NEW ZEALAND, NIGERIA, NORWAY, OMAN, PAKISTAN, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIAN FEDERATION, RWANDA, SAUDI ARABIA, SENEGAL, SERBIA, SEYCHELLES, SINGAPORE, SLOVAKIA (Slovak Republic), SLOVENIA, SOUTH AFRICA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIAN ARAB REPUBLIC, TAIWAN, United Republic of TANZANIA, THAILAND, TRINIDAD AND TOBAGO, TUNISIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, UZBEKISTAN, VIETNAM and ZIMBABWE.
  • 제품 설명
    Philips HeartStart XL+ Defibrillator/Monitor || Product Usage: || The HeartStart XL+ is intended for use in a hospital setting by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced life support or defibrillation. When operating as a semi-automatic external defibrillator in AED Mode, the HeartStart XL+ is suitable for use by medical personnel trained in basic life support that includes the use of an AED. When operating in Monitor, Manual Defibrillation or Pacing modes, the HeartStart XL+ is suitable for use by healthcare professionals trained in advance life support.
  • Manufacturer
  • 1 Event

제조사

  • 제조사 주소
    Philips Electronics North America Corporation, 3000 Minuteman Rd, Andover MA 01810-1032
  • 제조사 모회사 (2017)
  • Source
    USFDA

유사한 이름을 가진 의료기기 1건 유사한 이름을 가진 의료기기 11 건

데이터에 대해 더 자세히 알아보기 여기

  • 모델명 / 제조번호(시리얼번호)
    Manufactured between September 2002 and February 2013. The year of manufacture can be identified by the 2nd and 3rd characters in the serial number on the back of the AED in the following ranges:  Home/Onsite: A02x-xxxxx through A13B-xxxxx; FRx: B04x-xxxxx through B13B-xxxxx  NOTE: If the device was manufactured in 2013 and the 4th digit is the letter "C" or later (D, E, F...), it is not covered by the recall (i.e. A13G-02375).  In some cases, a device within the date range covered by the notification did not contain a resistor previously associated with a failure. In that case, a notification was not sent. Confirmation as to whether the device contains the resistor at issue can be done by contacting the firm.
  • 의료기기 분류등급
  • 의료기기 등급
    3
  • 이식된 장치?
    No
  • 유통
    Distribution was nationwide and to Puerto Rico. There was also government/military distribution. Foreign distribution was made to the following countries: Canada, Afghanistan, Algeria, Argentina, Aruba, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Botswana, Brazil, Brunei Darussalam, Cambodia, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Fiji, Finland, France, Georgia, Germany, Ghana, Greece, Guatemala, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macao, Malaysia, Malta, Mauritius, Mexico, Morocco, Namibia, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Niger, Norway, Oman, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Suriname, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirate, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Yemen, Zambia, and Zimbabwe.
  • 제품 설명
    Philips HeartStart FRx AED.
  • Manufacturer
  • 모델명 / 제조번호(시리얼번호)
    Manufactured between September 2002 and February 2013. The year of manufacture can be identified by the 2nd and 3rd characters in the serial number on the back of the AED in the following ranges:  Home/Onsite: A02x-xxxxx through A13B-xxxxx; FRx: B04x-xxxxx through B13B-xxxxx  NOTE: If the device was manufactured in 2013 and the 4th digit is the letter "C" or later (D, E, F...), it is not covered by the recall (i.e. A13G-02375).  In some cases, a device within the date range covered by the notification did not contain a resistor previously associated with a failure. In that case, a notification was not sent. Confirmation as to whether the device contains the resistor at issue can be done by contacting the firm.
  • 의료기기 분류등급
  • 의료기기 등급
    3
  • 이식된 장치?
    No
  • 유통
    Distribution was nationwide and to Puerto Rico. There was also government/military distribution. Foreign distribution was made to the following countries: Canada, Afghanistan, Algeria, Argentina, Aruba, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Botswana, Brazil, Brunei Darussalam, Cambodia, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Fiji, Finland, France, Georgia, Germany, Ghana, Greece, Guatemala, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macao, Malaysia, Malta, Mauritius, Mexico, Morocco, Namibia, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Niger, Norway, Oman, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Suriname, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirate, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Yemen, Zambia, and Zimbabwe.
  • 제품 설명
    Philips HeartStart (HS1) Onsite/Home AED
  • Manufacturer
  • 모델명 / 제조번호(시리얼번호)
    Philips HeartStart MRx monitor/defibrillators manufactured prior to June 2012 and used in external transport and Emergency Medical Service (EMS) environments, distributed worldwide
  • 의료기기 분류등급
  • 의료기기 등급
    3
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution - USA (nationwide) and the countries of : AFGHANISTAN ALBANIA ARGENTINA ARUBA AUSTRALIA AUSTRIA AZERBAIJAN BAHRAIN BANGLADESH BELGIUM BOLIVIA BOSNIA AND HERZEGOWINA BRAZIL BRUNEI DARUSSALAM BULGARIA CAMBODIA CANADA CHILE CHINA COLOMBIA COSTA RICA CROATIA (local name: Hrvatska) CYPRUS CZECH REPUBLIC DENMARK ECUADOR EGYPT EL SALVADOR ESTONIA ETHIOPIA FINLAND FRANCE GERMANY GREECE GUADELOUPE HONG KONG ICELAND INDIA INDONESIA IRAQ IRELAND ISRAEL ITALY JAPAN JORDAN KENYA KOREA, REPUBLIC OF KUWAIT LAOS LATVIA LEBANON LIBYAN ARAB JAMAHIRIYA LITHUANIA LUXEMBOURG MACAU MACEDONIA MALAYSIA MALTA MEXICO MOROCCO NAMIBIA NETHERLANDS NETHERLANDS ANTILLES NEW CALEDONIA NEW ZEALAND NICARAGUA NIGERIA NORWAY OMAN PAKISTAN PANAMA PAPUA NEW GUINEA PARAGUAY PERU PHILIPPINES POLAND PORTUGAL PUERTO RICO QATAR REUNION ROMANIA RUSSIAN FEDERATION SAUDI ARABIA SINGAPORE SLOVENIA SOUTH AFRICA SPAIN SRI LANKA SWEDEN SWITZERLAND TAIWAN, PROVINCE OF CHINA TANZANIA, UNITED REPUBLIC OF THAILAND TRINIDAD AND TOBAGO TUNISIA TURKEY UGANDA UKRAINE UNITED ARAB EMIRATES UNITED KINGDOM UZBEKISTAN VENEZUELA VIET NAM YEMEN
  • 제품 설명
    Philips HeartStart MRx monitor/defibrillators || Models: M3535A and M3536A, M3536J, M3536M, M3536MC, M3536M1, M3536M2, M3536M3, M3536M4, M3536M5, M3536M6 || The Heartstart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician
  • Manufacturer
  • 모델명 / 제조번호(시리얼번호)
    Manufactured between the dates of July 12, 2011 through September 2, 2011 and January 19, 2012 through May 22, 2012.  Serial numbers within the ranges of: US00550047 to US00551654 and US00556357 to US00559493 respectively.   There are also some additional MRx units that were repaired and may contain compromised components. These MRx unit serial numbers are identified below: US00210180, US00210183, US00210738, US00211000, US00213281 US00214706, US00316827, US00318163, US00319241, US00328491 US00329811, US00333101, US00536832, US00538140, US00543161 US00543825, US00544081, US00544214, US00546107, US00548046
  • 의료기기 분류등급
  • 의료기기 등급
    3
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution - US (nationwide) and the countries of: AUSTRALIA, AUSTRIA, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHINA, COLOMBIA, EL SALVADOR, ETHIOPIA, FRANCE, GERMANY, HONG KONG, INDIA, INDONESIA, ISRAEL, ITALY, JAPAN, JORDAN, KENYA, KOREA, REPUBLIC OF, LEBANON, LIBYAN ARAB JAMAHIRIYA, MEXICO, MOROCCO, NETHERLANDS, NETHERLANDS ANTILLES, NICARAGUA, NORWAY, PERU, PHILIPPINES, POLAND, QATAR, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, THAILAND, TRINIDAD AND TOBAGO, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA, and VIET NAM.
  • 제품 설명
    Philips HeartStart MRx monitor/defibrillators || Models: M3535A, M3536A, M3536M4, M3536M5, and M3536M6 || Product Usage: || The Heartstart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician
  • Manufacturer
  • 모델명 / 제조번호(시리얼번호)
    1)Philips brand FRx Model Number 861304 under 510(k) K050004, sold nationally and Internationally, Serial numbers range from BO4L-00001 through B12F-00286 and represents 181,226 devices; 2) Laerdal brand FRx Model Number 861305 unfrt 510(k) K050004, sold nationally and Internationally, serial numbers range from B04L-00013 through B12E-04135 and represents 20,372 devices.  Please note the serial number range represents first and last serial number for each model. Serial numbers are not continguous.
  • 의료기기 분류등급
  • 의료기기 등급
    3
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution - USA (nationwide) and the countries of Japan and Malaysia.
  • 제품 설명
    Philips HeartStart FRx 861304 Automated External Defibrillator Manufactured by Philips Medical Systems, Seattle, WA 98121 USA; and Laerdal branded FRx 861305 || The FRx Defibrillator is intended to treat ventricular fibrillation, the most common cause of sudden cardiac arrest. Using voice prompts, light emitting diodes (LEDs) and buttons, the user is guided through the response. The FRx uses a SMART biphasic, impedance compensating exponential waveform to deliver a nominal 150 J to adults and nominal 50 J to infants/children.
  • Manufacturer
하나 더 6 건 더