Device Recall Oncor AvantGarde Medical Linear Accelerator

  • 모델명 / 제조번호(시리얼번호)
    Serial Numbers: 3576 3938 70-4283 3738 3817 3987 3988 4095 70-4368 4016 3904 3926 4079 4082 70-4077 70-4113 70-4197 3905 3934 3852 3737 3758 3856 3891 3929 3937 70-4163 3895 3896 4049 70-4201 70-4184 5038 70-4213 70-4223 5012 70-4311 3932 3739 3806 3884 3912 3941 70-4297 70-4314 and 70-4217
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide-USA including states of AL, AK, AZ, AR, CA, CO, DE, DC, FL, GA, IL, IN, KS, KY, LA, MD, ME, MI, MN, MS, MO, MN, NE, NV, NH, NJ, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, VT, VA, WA, WV, WI, and WY and countries of Angola, Australia, Austria,Belgium,Brazil,Canada,Chile,Colombia,Costa Rica, Croatia,Cyprus,The Czech Republic,Denmark,DPR of Korea,El Salvador,France,Germany,Greece,Hungary,India,Iran,Ireland,Italy,Japan,Jordan,Malaysia,Mexico, The Netherlands,New Zealand,Norway,P.R. China,Pakistan,Peru,Philippines,Poland,Portugal,Republic of Korea,Romania,Russian Fed.Saudi Arabia,Singapore,Slovakia,South Africa,Spain,Sweden,Taiwan,Thailand,Trinidad,Tobago,Turkey, U.A.E.,Ukraine, The United Kingdom, and Venezuela
  • 제품 설명
    Oncor Avante-Garde Medical Linear Accelerator, Material Number: 5857920, with control console software versions 6.5/7.2, 8.0, 9.0 and 9.1, Manufactured by Siemens Medical Solutions USA, Inc, Concord, CA
  • Manufacturer
  • 1 Event

제조사

  • 제조사 주소
    Siemens Medical Solutions USA, Inc, 4040 Nelson Ave, Concord CA 94520-1200
  • Source
    USFDA