Carina Ventilator 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Draeger Medical, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    80027
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1954-2018
  • 사례 시작날짜
    2018-04-26
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Ventilator, continuous, facility use - Product Code CBK
  • 원인
    The valve assembly material may become corroded due to insufficient cleaning by a supplier. there is a risk of the breathing gas becoming contaminated in the affected devices.
  • 조치
    On April 27, 2018, the recalling firm contacted its consignees by telephone prior to sending out Urgent Medical Device Recall notifications via e-mail. Customers were instructed to remove affected devices from operation immediately. The firm provided replacement devices so customers could send their affected ventilator to the firm's repair center for update once the required parts were available. Customers were also asked to inform all affected users at each site. If you have any questions regarding this letter please contact Michael Kelhart between the hours of 8:00 AM  4:30 PM EST at 1-800-437-2437. For questions regarding the operation and/or servicing of your Carina Ventilator please contact Dr¿ger Service Technical Support between the hours of 8:00 AM  8:00 PM EST at 1-800-437-2437.

Device

  • 모델명 / 제조번호(시리얼번호)
    Serial nos. ASKM-0065 ASKM-0066 ASKM-0106 ASKM-0134 ASKM-0137 ASKM-0150 ASKM-0151 ASKN-0009
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Distributed to the states of CT, IL, MN, NV, SD, and WA.
  • 제품 설명
    Carina Ventilator, ventilator. Catalog no. 5704110. || Long-term ventilator for treatment in hospitals by a qualified medical personnel.