Depuy P.F.C. E Knee System 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Depuy Orthopaedics, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    44893
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0403-2008
  • 사례 출판 날짜
    2007-12-06
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2008-04-30
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Tibial Component - Product Code NJL
  • 원인
    An incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant.
  • 조치
    Consignee hospitals and implanting physicians were notified of the recall via an Urgent Information-Recall Notice letter and Dear Doctor letters dated 9/27/07. The hospitals were instructed to determine whether they had any of the recalled products and to contact their sales representative to pick up the product. Depuy sales representatives were notified to collect the products from consignee hospitals via email on 10/1/07. The decision to discuss this product defect with patients who had already been implanted was left to the implanting physicians.

Device

  • 모델명 / 제조번호(시리얼번호)
    Lots 77471A, 81083A and 81271A.
  • 의료기기 분류등급
  • 의료기기 등급
    3
  • 이식된 장치?
    Yes
  • 유통
    Nationwide.
  • 제품 설명
    Depuy P.F.C. ¿ E Knee System All Polyethylene Curved Tibial Component, 3, 8 mm, 71mm M/L, 47 mm A/P, UHMWPE, sterile; Product Code/ REF 96-0640.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Depuy Orthopaedics, Inc., 700 Orthopaedic Dr, Warsaw IN 46582-3994
  • Source
    USFDA