Device Recall Beckman Coulter Cytomics FC 500 Flow Cytometry System 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Beckman Coulter Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    35643
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1259-06
  • 사례 시작날짜
    2006-05-22
  • 사례 출판 날짜
    2006-07-26
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2012-04-26
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Flow Cytometry System - Product Code LXG
  • 원인
    Beckman coulter has determined that under certain conditions a system error generated during a pause and rotate routine may result in a sample misidentification with a risk of reporting erroneous results.
  • 조치
    A Product Corrective Action letter was sent via US mail the week of May 22, 2006 to all FC500 customers. They were informed that under certain conditions a system error generated during a Pause and Rotate routine may result in a sample misidentification with a risk of reporting erroneous results. The letter includes the sequence in which the issue occurrs: 1) During the acquistion of a tube, the Pause and Rotate Button is selected. 2) A system error is generated during the Pause and Rotate routine (The error message table may be viewed by going to the CXP Help Menu, select the CXP Help Menu Item, then to Section 13.5). 3) An error message appears in the Cytometer Status, an audible alarm sounds, the run aborts, the carousel ejects, and the system automatically transitions from Verification to Run Initialization and then to Awaiting Sample 4) The Worklist is continued or a new Worklist is loaded and started. 5) After the Worklist is continued, or loaded and started, during the acquistion of a tube, the Pause and Rotate Button is selected. 6) The carousel Pauses but does not Rotate. 7) When the Start/Continue or Restart button is selected, the carousel may move to the wrong tube and aspirate sample resulting in a sample misidentification. If no system errors are generated during a Pause and Rotate routine, the system will function correctly. The letter also provides steps for immediate user preventive action if a system error is generated during the Pause and Rotate Routine to remedy the issue: 1) Exit CXP Software. 2) Power off the Cytometer using the FC OFF utility 3) Turn the Cytometer back on. 4) Recovery should be complete and the laboratory may resume normal usage of the system. Customers are asked to share the information with their laboratory staff and to retain this notification as part of their laboratory Quality System documentation and to complete and return the enclosed Fax Response Form for the firm''s records.

Device

  • 모델명 / 제조번호(시리얼번호)
    Software Versions 2.0 and 2.1
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    Nationwide and Canada
  • 제품 설명
    Cytomics FC 500 Flow Cytometry System Part Numbers 626553 (TN; FC500 with UPS), 6605627 (TN; 5 CLR, FC500 (100V)), 6605628 (TN; 5 CLR, FC500 (120V)) , 6605629 (TN; 5 CLR, FC500 (220V)), 6605630 (TN; 5 CLR, FC500 (240V)) with CXP Software Versions 2.0 and 2.1
  • Manufacturer

Manufacturer

  • 제조사 주소
    Beckman Coulter Inc, 200 S Kraemer Blvd, Brea CA 92822-6208
  • Source
    USFDA