The temperature of the incubation lids of the minc bench top incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.
조치
Cook plans to begin notifying consignees of the recall on January 12, 2018
Written recall communication letters will be mailed via courier to US consignees
Nationwide (K-IMNC-100 -US G46022) . (K-IMNC-100 G20079 manufactured in Australia and distributed worldwide)
제품 설명
MINC Benchtop Incubator (Mini Incubator) || catalog (K-MINC-1000) G20079 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture.