Device Recall Consult Diagnostics Eon Calcium Reagent For InVitro Diagnostic Use 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Vital Diagnostics, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    66202
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0733-2014
  • 사례 시작날짜
    2013-08-29
  • 사례 출판 날짜
    2014-01-14
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2015-08-26
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Azo dye, calcium - Product Code CJY
  • 원인
    High positive calcium bias on plasma sample versus results on serum.
  • 조치
    Vital Diagnostics issued a Recall Notice dated August 29, 2013, to all affected customers via certified mail. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed not to use plasma for the determination of total calcium when using the affected product. The use of serum is acceptable, thus the product can still be used to determine total calcium with this sample type. Vital Diagnostics released a corection to the package insert for the Envoy Calcium Reagent. Customers with the affected product in their laboratory, were instructed to discard the insert packed with the product and replace it with the insert enclosed with this bulletin. If product was transferred to another location, the bulletin should be forwarded to that location. The package insert was revised indicating to only use Serum samples when running Calcium. Vital Diagnostics issued a revised (expanded) notice dated September 19, 2013, to include the Consult Diagnostics¿ Eon Calcium Reagent (PN: 1429) requesting users to discontinue using the single vial Eon Calcium kit (PN: 1429). Customers with questions were instructed to call Vital Diagnostics Technical Support at 1-855-354-8324.. For questions regarding this recall call 401-642-8400.

Device

  • 모델명 / 제조번호(시리얼번호)
    All lots
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Nationwide Distribution
  • 제품 설명
    Consult Diagnostics¿ Eon Calcium Reagent For In-Vitro Diagnostic Use || PN: 1429 || For the quantitative measurement of calcium in plasma and serum