Device Recall Cook Vacuum Pump 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Cook Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    79713
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1461-2018
  • 사례 시작날짜
    2018-03-22
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Accessory, assisted reproduction - Product Code MQG
  • 원인
    Pump was found to be constructed from crimp connectors that do not meet the fixation wiring requirements and can cause failure of the device to operate or electric shock or burn to the user.
  • 조치
    An Urgent: Medical Device Correction letter, dated March 20, 2018, was sent to consignees. The letter identified the affected device and provided a description of the problem. The firm stated that it would be replacing the mains wiring for each unit. An authorized service agent will contact customers to arrange for the affected devices to be corrected. Customers are to follow the instructions regarding the residual current device (RCD) provided in the letter, while they are waiting for their devices to be corrected. Additionally, customers should complete the Acknowledgement and Receipt Form within 5 business days of receiving the letter. Questions or concerns should be directed to Cook Medical Capital Equipment Service at 855-207-7214.

Device

  • 모델명 / 제조번호(시리얼번호)
    All lots
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution -- USA, Argentina, Bolivia, Brazil, Canada, Colombia, Guatemala, Mexico, Panama, Peru, Uruguay, and Venezuela.
  • 제품 설명
    Cook Vacuum Pump ( Cook Aspiration Unit) K-MAR-5200-US, GPN G49275. || Intended for the aspiration of body fluids and cells, in particular oocyte aspiration.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Cook Inc., 750 N Daniels Way, Bloomington IN 47404-9120
  • 제조사 모회사 (2017)
  • Source
    USFDA