System, Test, Vitamin D - Product Code MRG
원인
An internal inspection indicates that calibrator 2 and conjugate are switched in their positions in the integral. this will result in calibrator 1 rlu's similar to that of background rlu, and calibrator 2 rlu's near expected range. a curve can not be calculated from this data.
조치
DiaSorin Customer Notification dated 02/14/05 was sent to customers instructing them to destroy all kits from the affected lot number. Customers were also asked to complete and return a form indicating how many kits were destroyed.