U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
데이터 추가 비고
Antigens, All Types, Escherichia Coli - Product Code GMZ
원인
A potential contamination was noted in the conjugate in the prospect shiga toxin e. coli microplate assay. the particulate in the conjugate may cause the dropper tip to become plugged.
조치
Remel contacted each customer with a letter, by mail, requesting a count of the affected lots to be returned. Product would be replaced upon customer request.
ProSpecT Shiga Toxin E. coli (STEC) Micorplate Assay. Catalog #: 2474048 (48 well plate) and 2474096 (96 well plate). For qualitative detection of Shiga toxins (STX1 & STX 2) in aqueous extracts of fecal specimens and fecal enriched broth cultures.