Device Recall SMART PERFUSION PACK, LONG PACK 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Sorin Group USA, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    77810
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-2783-2017
  • 사례 시작날짜
    2017-06-29
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Tubing, pump, cardiopulmonary bypass - Product Code DWE
  • 원인
    Medtronic informed livanova that some of their dlp pressure disposable display sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). the result is a potential sterile barrier breach in the display sets. these display sets are included with livanova perfusion tube sets (pts) which were then distributed to customers.
  • 조치
    The firm initiated their recall verbally on 06/29/2017 and following with letters mailed on 07/13/2017. The actions to be taken by the customer: 1. Notify staff to not use the Medtronic labeled device contained within the LivaNova Perfusion Tube Set packs. 2. If the set is a standalone component within the pack, remove the device upon opening the pack at the time of the procedure. 3. If the device is connected to the circuit, remove the device and attach a new unit upon opening the pack. 4. Identify the Medtronic device with one of the barcode pack labels attached to the lidstock of the LivaNova perfusion pack. 5. Collect the devices for return to LivaNova until all affected perfusion sets have been consumed. In order to assist in controlling inventory, the LivaNova sales representative will have labels available to identify packs containing the affected devices. 6. Contact LivaNova Customer Support at (800) 650-2623 (M-F 8AM to 5PM MT) to arrange for the return of the affected product, and to order equivalent replacement product.

Device

  • 모델명 / 제조번호(시리얼번호)
    Lot Numbers: 1625200040, 1635400048, 1704700039, 1706700087, 1711400055, 1717200038
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Nationwide
  • 제품 설명
    SORIN GROUP, SMART PERFUSION PACK, LONG PACK, REF 088512600, STERILE EO, Rx Only, 1 EA
  • Manufacturer

Manufacturer

  • 제조사 주소
    Sorin Group USA, Inc., 14401 W 65th Way, Arvada CO 80004-3503
  • 제조사 모회사 (2017)
  • Source
    USFDA