Device Recall TSX101: Aquilion 32/64 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Toshiba American Medical Systems Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    59519
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-3221-2011
  • 사례 시작날짜
    2010-12-08
  • 사례 출판 날짜
    2011-09-16
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2013-03-18
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    System, x-ray, tomography, computed - Product Code JAK
  • 원인
    There is the potential that the current may not be controlled correctly in examinations using the ecg modulation function, which adjusts the x-ray output in synchronization with the ecg waveform.
  • 조치
    Toshiba America Medical Systems (TAMS) issued an Urgent: Medical Device Correction notification to consignees via USPS return receipt mail. The letter identified the affected product, the problem, and corrective action. The firm plans to install revised firmware to prevent the occurrence of the problem. A Toshiba Service Representative will contact customers to schedule an appointment for installation. The firm asked customers to stop using the Power Reduction option during ECG modulated scanning until the new firmware has been installed. Customers are to share the information in the letter with all users, reviewing radiologist, and the clinical engineering or Biomedical group at their facility. Customers are also to complete and return the attached reply form via fax or e-mail to TAMS as proof of notification. Questions pertaining to the letter should be directed to 800-421-1968 or customers' local Toshiba Representative at 800-521-1968.

Device

  • 모델명 / 제조번호(시리얼번호)
    Units manufactured from January 2005 to September 2007
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Nationwide Distribution.
  • 제품 설명
    TSX-101: Aquilion 32/64; System, X-Ray, Tomography, Computed.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Toshiba American Medical Systems Inc, 2441 Michelle Dr, P.O. Box 2068, Tustin CA 92781-2068
  • 제조사 모회사 (2017)
  • Source
    USFDA